RecallWatchUSA

Home / Recalls / Medical Devices

GE Medical Systems, LLC recalls GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed

Recalled March 19, 2025 · FDA recall Z-1297-2025 · GE Medical Systems

Hazard
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls. · FDA Class II
Units
56 units
Sold at
Worldwide distribution.

Description

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

Product
GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed
Manufacturer(s)
GE Medical Systems, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

GE Medical Systems, LLC recalls GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed | RecallWatchUSA