RecallWatchUSA

Home / Recalls / Medical Devices

GE Medical Systems, LLC recalls Optima XR240amx, Mobile Digital Radiographic System

Recalled July 22, 2026 · FDA recall Z-2728-2026 · GE Medical Systems

Hazard
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam. · FDA Class II
Units
24 units
Sold at
Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.

Description

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

Product
Optima XR240amx, Mobile Digital Radiographic System
Manufacturer(s)
GE Medical Systems, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov