RecallWatchUSA

Home / Recalls / Medical Devices

GE Medical Systems, LLC recalls Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile

Recalled January 22, 2025 · FDA recall Z-0918-2025 · GE Medical Systems

Hazard
There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time. · FDA Class II
Units
7 systems
Sold at
Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK.

Description

There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.

Product
Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facility infrastructure and clinical information systems, Model CSSXB
Manufacturer(s)
GE Medical Systems, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

GE Medical Systems, LLC recalls Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile | RecallWatchUSA