GE OEC Medical Systems, Inc recalls GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedure

Recalled January 21, 2015 · FDA recall Z-0841-2015 · GE OEC Medical Systems

Hazard
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3). · FDA Class II
Units
61 units installed in US
Sold at
Worldwide Distribution

Description

mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).

Product
GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
Manufacturer(s)
GE OEC Medical Systems, Inc