Gen-Probe Inc recalls APTIMA Urine Collection Kit; Catalog Number 301040, Lots 27231, 27232, 27233, and 27234. Product Usage: The APTIMA
Recalled June 25, 2014 · FDA recall Z-1793-2014 · Gen-Probe
Hazard
Incorrect labeling was used on each pouch of the affected lots of APTIMA Urine Specimen Collection Kits. The outer box of the kits are identified correctly as "APTIMA Urine Collection Kits"; however, the inner pouches are incorrectly labeled as "APTIMA Unisex Swab Collection Kit for Endocervical and Male Urethral Swab Specimens." · FDA Class III
Units
4,740 units
Sold at
Worldwide Distribution - US Nationwide and the countries of Canada (CA), Denmark (DK), France (FR), Germany (DE), and the United States (US), Belgium (BE), United Kingdom (GB).
Description
Incorrect labeling was used on each pouch of the affected lots of APTIMA Urine Specimen Collection Kits. The outer box of the kits are identified correctly as "APTIMA Urine Collection Kits"; however, the inner pouches are incorrectly labeled as "APTIMA Unisex Swab Collection Kit for Endocervical and Male Urethral Swab Specimens."
Product
APTIMA Urine Collection Kit; Catalog Number 301040, Lots 27231, 27232, 27233, and 27234. Product Usage: The APTIMA Urine Specimen Collection Kit for Male and Female Urine Specimens is for use with APTIMA assays. The APTIMA Urine Specimen Collection Kit is intended to be used for the collection and transport of male or female urine specimens.
Manufacturer(s)
Gen-Probe Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →