Gendex Corp recalls eXpertDC intraoral x-ray, models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN, 55 Inch rea

Recalled October 8, 2014 · FDA recall Z-0009-2015 · Gendex

Hazard
Two (2) failures of the spring link components within the articulated arm assembly. The spring links broke during normal assembly process for the articulated arm during first cycle of the arm, prior to the 14lb tube head (full load) being installed onto the assembled articulated arm. · FDA Class II
Units
651
Sold at
Worldwide Distribution - US Nationwide and the countries of Europe, Canada, Australia, China, Russia, Taiwan, Singapore, and Thailand.

Description

Two (2) failures of the spring link components within the articulated arm assembly. The spring links broke during normal assembly process for the articulated arm during first cycle of the arm, prior to the 14lb tube head (full load) being installed onto the assembled articulated arm.

Product
eXpertDC intraoral x-ray, models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN, 55 Inch reach; eXpert DC, WW, 75 Inch reach; eXpert DC, WW, 65 Inch reach; eXpert DC, WW, 55 Inch reach; OEM Arm Assembly w/o cables; Assy, Articulated Arm & Tubehead, eDC; and Assy, Articulated Arm & Tubehead, 765DC Product Usage: intraoral x-ray
Manufacturer(s)
Gendex Corp

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov