Home / Recalls / Medical Devices
Getinge Usa Sales Inc recalls Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Fl
Recalled February 19, 2025 · FDA recall Z-1074-2025 · Getinge Usa Sales
Hazard
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia. · FDA Class I
Units
6 units
Sold at
Domestic distribution to Kentucky and Connecticut.
Description
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
Product
Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.
Manufacturer(s)
Getinge Usa Sales Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →