RecallWatchUSA

Home / Recalls / Medical Devices

Globus Medical, Inc. recalls ALTERA Spacer ALTERA Spacer, 10x36, 10-14mm, 15¿

Recalled September 2, 2020 · FDA recall Z-2870-2020 · Globus Medical

Hazard
Specific lots of ALTERA¿ implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy. · FDA Class II
Units
N/A
Sold at
Worldwide Distribution

Description

Specific lots of ALTERA¿ implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.

Product
ALTERA Spacer ALTERA Spacer, 10x36, 10-14mm, 15¿
Manufacturer(s)
Globus Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov