Home / Recalls / Medical Devices
GN Otometrics recalls Gold Cup Leadwire and Snap Leadwire with the following component descriptions: 1. 24" Snap Leadwire, Blue; 2. 4
Recalled October 2, 2019 · FDA recall Z-2553-2019 · GN Otometrics
Hazard
The electrode lead connector has a protruding conductor beyond insulation of approximately 0.1 mm. · FDA Class II
Units
4,882 units (41950 individual electrode leads supplied In 5,7 & 10 packs)
Sold at
Worldwide distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Cote D'ivoire, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, France, Germany, Greece, Guyana, Hong Kong, Hungary, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Korea, Korea, Republic of, Kuwait, Lebanon, Luxembourg, Macedonia, The Former Yugoslav, Malaysia, Maldives, Mayotte, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, Zimbabwe.
Description
The electrode lead connector has a protruding conductor beyond insulation of approximately 0.1 mm.
Product
Gold Cup Leadwire and Snap Leadwire with the following component descriptions: 1. 24" Snap Leadwire, Blue; 2. 40" Snap Leadwire, Blue; 3. 24" Snap Leadwire, Red; 4. 40" Snap Leadwire, Red; 5. 24" Snap Leadwire, Green; 6. 40" Snap Leadwire, Green; 7. 24" Snap Leadwire, Brown; 8. 40" Snap Leadwire, Brown; 9. 24" Snap Leadwire, Orange; 10. 40" Snap Leadwire, Orange; 11. 24" Snap Leadwire, White; 12. 40" Snap Leadwire, White; 13. 24" Snap Leadwire, Black; 14. 40" Snap Leadwire, Black; 15. 24" Snap Leadwire, Yellow; 16. 40" Snap Leadwire, Yellow; 17. 24" Snap Leadwire, Set of 5; 18. 24" Snap Leadwire, Set of 7; 19. 40" Snap Leadwire, Set of 5; 20. 40" Snap Leadwire, Set of 7; 21. 24" Snap Leadwire, Set of 5; 22. 40" Snap Leadwire, Set of 5; 23. 48" Gold Cup Leadwires, Blue; 24. 48" Gold Cup Leadwires, Red; 25. 48" Gold Cup Leadwires, Green; 26. 48" Gold Cup Leadwires, Brown; 27. 48" Gold Cup Leadwires, Orange; 28. 48" Gold Cup Leadwires, White; 29. 48" Gold Cup Leadwires, Black; 30. 48" Gold Cup Leadwires, Yellow; 31. 48" Gold Cup Leadwires, Set of 5; 32. 48" Gold Cup Leadwires, Set of 8; Product Usage: The snap electrode leads are used in conjunction with disposable electrodes and the Gold Cup electrodes are reusable electrodes and leads. Both types of electrodes and leads are used for evoked potential testing for audiological, neurological and/or vestibular testing. The electrode leads are connected to the main device by the electrode lead connector and the electrode is attached to the patient contact end of electrode lead.
Manufacturer(s)
GN Otometrics
Made in
Denmark
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →