Graphic Controls Acquisition Corporation recalls CLARAVUE Pre-wired Electrode - IP SET M3 RT ADULT-US Model: 32028777 50600 US
Recalled April 20, 2022 · FDA recall Z-0936-2022 · Graphic Controls Acquisition
Hazard
Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation. · FDA Class II
Units
560 units
Sold at
US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT.
Description
Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.
Product
CLARAVUE Pre-wired Electrode - IP SET M3 RT ADULT-US Model: 32028777 50600 US
Manufacturer(s)
Graphic Controls Acquisition Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →