Home / Recalls / Medical Devices
Greiner Bio-One North America, Inc. recalls 3mL 9NC Coagulation Sodium Citrate 3.2%, Item #454334, Greiner Bio-One North America Inc., 4238 Capital Drive, Monroe,
Recalled August 15, 2012 · FDA recall Z-2154-2012 · Greiner Bio-One North America
Hazard
Part of the lot received a lower concentration of citrate solution which may cause falsely lower values for common coagulation tests. · FDA Class II
Units
838.9 cases (1,006,700 pieces)
Sold at
Nationwide Distribution
Description
Part of the lot received a lower concentration of citrate solution which may cause falsely lower values for common coagulation tests.
Product
3mL 9NC Coagulation Sodium Citrate 3.2%, Item #454334, Greiner Bio-One North America Inc., 4238 Capital Drive, Monroe, NC 28110. Intended Use: Venous blood collection tubes.
Manufacturer(s)
Greiner Bio-One North America, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →