Greiner Bio-One North America, Inc. recalls VACUETTE Safety Blood Collection Set + Luer Adapter, 23x12" Safety Blood Collect LA Product, Item number 450096. Usage:
Recalled February 27, 2013 · FDA recall Z-0835-2013 · Greiner Bio-One North America
Hazard
The VACUETTE Safety Blood Collection Set + Luer Adapter may display leaking (small droplets) during use due to a crack on the luer connector piece. · FDA Class II
Units
192,000 pieces (192 cases)
Sold at
US Nationwide Distribution in the states of FL, IL, KY, NY, NC, PA, OH, TX, VA
Description
The VACUETTE Safety Blood Collection Set + Luer Adapter may display leaking (small droplets) during use due to a crack on the luer connector piece.
Product
VACUETTE Safety Blood Collection Set + Luer Adapter, 23x12" Safety Blood Collect LA Product, Item number 450096. Usage: The Safety Blood Collection Set is used in routine venipuncture procedures.
Manufacturer(s)
Greiner Bio-One North America, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →