RecallWatchUSA

Home / Recalls / Medical Devices

Haemonetics Corporation recalls JAPAN CS ELITE FASTPACK, 225ML - Product Usage: its related accessory components are intended for use to recover blood s

Recalled February 5, 2020 · FDA recall Z-0902-2020 · Haemonetics

Hazard
Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing. · FDA Class II
Units
8114
Sold at
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.

Description

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

Product
JAPAN CS ELITE FASTPACK, 225ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.
Manufacturer(s)
Haemonetics Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Haemonetics Corporation recalls JAPAN CS ELITE FASTPACK, 225ML - Product Usage: its related accessory components are intended for use to recover blood s | RecallWatchUSA