RecallWatchUSA

Home / Recalls / Medical Devices

Haemonetics Corporation recalls TEG Manager software versions used in conjunction with TEG 5000 are 1.1.0, 3.0.0, 4.0.0, 4.1.0 and 4.1.1. Intended: Fo

Recalled April 8, 2020 · FDA recall Z-1617-2020 · Haemonetics

Hazard
Software defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager test result screen are rounded to the nearest whole number and lead to TEG Manager displaying an out of range alert when the test result is actually in range, or vice versa. · FDA Class II
Units
88 units
Sold at
US Nationwide distribution.

Description

Software defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager test result screen are rounded to the nearest whole number and lead to TEG Manager displaying an out of range alert when the test result is actually in range, or vice versa.

Product
TEG Manager software versions used in conjunction with TEG 5000 are 1.1.0, 3.0.0, 4.0.0, 4.1.0 and 4.1.1. Intended: For in vitro coagulation studies
Manufacturer(s)
Haemonetics Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Haemonetics Corporation recalls TEG Manager software versions used in conjunction with TEG 5000 are 1.1.0, 3.0.0, 4.0.0, 4.1.0 and 4.1.1. Intended: Fo | RecallWatchUSA