Halyard Health, Inc recalls NovaPlus Trach Care* Closed Suction System for Adults, Clear T-Piece MDI Adapter, 14F (Product Code 2V220158) A trach
Recalled September 9, 2015 · FDA recall Z-2501-2015 · Halyard Health
Hazard
The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position. · FDA Class II
Units
N/A
Sold at
Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada
Description
The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.
Product
NovaPlus Trach Care* Closed Suction System for Adults, Clear T-Piece MDI Adapter, 14F (Product Code 2V220158) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.
Manufacturer(s)
Halyard Health, Inc