RecallWatchUSA

Home / Recalls / Medical Devices

Hamilton Medical, Inc. recalls HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology: The HAMILTON-G5 ventilator is d

Recalled November 11, 2015 · FDA recall Z-0151-2016 · Hamilton Medical

Hazard
Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the operator; this situation can occur regardless of the selected patient group (neonatal, pediatric and adult). · FDA Class I
Units
1128
Sold at
Nationwide Distribution

Description

Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the operator; this situation can occur regardless of the selected patient group (neonatal, pediatric and adult).

Product
HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology: The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients
Manufacturer(s)
Hamilton Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Hamilton Medical, Inc. recalls HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology: The HAMILTON-G5 ventilator is d | RecallWatchUSA