RecallWatchUSA

Home / Recalls / Medical Devices

Hamilton Medical, Inc. recalls Hamilton-MR1, REF: 161010, Intensive Care Ventilator

Recalled September 27, 2023 · FDA recall Z-2528-2023 · Hamilton Medical

Hazard
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation. · FDA Class I
Units
12 units
Sold at
Distribution US nationwide, including Puerto Rico. OUS: Not provided

Description

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Product
Hamilton-MR1, REF: 161010, Intensive Care Ventilator
Manufacturer(s)
Hamilton Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov