Hamilton Medical, Inc. recalls Hamilton-MR1 Ventilator: Catalog# 161010 The Hamilton MR1 Ventilator is intended to provide pressure ventilator suppor

Recalled December 7, 2016 · FDA recall Z-0666-2017 · Hamilton Medical

Hazard
Oxygen tubing and the oxygen connector of the Hamilton-MR1 could potentially become loose during the preparation for ventilation. · FDA Class II
Units
44 units
Sold at
Nationwide Distribution

Description

Oxygen tubing and the oxygen connector of the Hamilton-MR1 could potentially become loose during the preparation for ventilation.

Product
Hamilton-MR1 Ventilator: Catalog# 161010 The Hamilton MR1 Ventilator is intended to provide pressure ventilator support to adults and pediatrics, and optionally infants and neonates. Intended areas of use include: MRI Dept., Intensive, intermediate, emergency and long-term acute care as well as transfer of patients within a hospital.
Manufacturer(s)
Hamilton Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov