Home / Recalls / Medical Devices
Hamilton Medical, Inc. recalls Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator
Recalled September 27, 2023 · FDA recall Z-2527-2023 · Hamilton Medical
Hazard
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation. · FDA Class I
Units
1216 units
Sold at
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Description
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Product
Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator
Manufacturer(s)
Hamilton Medical, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →