Hand Biomechanics Lab Inc recalls Agee-WristJack¿ Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture

Recalled October 26, 2016 · FDA recall Z-0136-2017 · Hand Biomechanics Lab

Hazard
A number of Agee WristJack¿ Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the surgeon experiencing difficulty fine-tuning the length (red) and radial/ulnar alignment (blue) adjustments on the device. · FDA Class II
Units
65 units
Sold at
Nationwide Distribution

Description

A number of Agee WristJack¿ Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the surgeon experiencing difficulty fine-tuning the length (red) and radial/ulnar alignment (blue) adjustments on the device.

Product
Agee-WristJack¿ Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. This is a single use system used for a trauma related surgery The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. The non-perishable sterile system is packaged in an inner then outer tray and shipped in a cardboard shelf box. Orthopedic:The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures.
Manufacturer(s)
Hand Biomechanics Lab Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Hand Biomechanics Lab Inc recalls Agee-WristJack¿ Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture | RecallWatchUSA