HEALTH ROBOTICS S.R.L. recalls i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System Device Listing Number
Recalled January 18, 2017 · FDA recall Z-0989-2017 · HEALTH ROBOTICS S.R.L
Hazard
The firm has initiated a product correction due to a bag carousel motor error causing misalignment and resulting in a mislabeling of final IV bag preparations. · FDA Class II
Units
121 systems
Sold at
Shipped throughout United States.
Description
The firm has initiated a product correction due to a bag carousel motor error causing misalignment and resulting in a mislabeling of final IV bag preparations.
Product
i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System Device Listing Number E454448
Manufacturer(s)
HEALTH ROBOTICS S.R.L.
Made in
Italy
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →