RecallWatchUSA

Home / Recalls / Medical Devices

Healthmark Industries Co., Inc. recalls ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)

Recalled June 11, 2025 · FDA recall Z-1910-2025 · Healthmark Industries

Hazard
During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose. · FDA Class II
Units
193035 units
Sold at
US Nationwide and Internationally to countries of: Canada, Malaysia.

Description

During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.

Product
ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)
Manufacturer(s)
Healthmark Industries Co., Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Healthmark Industries Co., Inc. recalls ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample) | RecallWatchUSA