Home / Recalls / Medical Devices
Heartsine Technologies, Limited recalls HeartSine Samaritan(R) 300/300P PAD devices with serial numbers 0400000501 to 0700032917; 08A00035000 to 10A0070753; 10C
Recalled November 7, 2012 · FDA recall Z-0124-2013 · Heartsine Technologies
Hazard
Certain Samaritan 300/300P PAD devices may experience one or both of the following conditions that could affect the ability to deliver therapy to a patient in an SCA event if needed: Issue 1 (On/Off Issue): the device may turn itself on and off without input from the user. When this occures, the normal sequence of audible instruction prompts that occurs when the device is turned on will be emit · FDA Class I
Units
81,982 in total
Sold at
Worldwide Distribution and US (nationwide) in the following states: AK, AL, AP, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, and WI.
Description
Certain Samaritan 300/300P PAD devices may experience one or both of the following conditions that could affect the ability to deliver therapy to a patient in an SCA event if needed: Issue 1 (On/Off Issue): the device may turn itself on and off without input from the user. When this occures, the normal sequence of audible instruction prompts that occurs when the device is turned on will be emit
Product
HeartSine Samaritan(R) 300/300P PAD devices with serial numbers 0400000501 to 0700032917; 08A00035000 to 10A0070753; 10C00200000 to 10C00210106 Product Usage: Samaritan 300/300P PAD is indicated for use on victims of sudden cardiac arrest who are exhibiting the following signs: Unconsciousness Not breathing Without circulation
Manufacturer(s)
Heartsine Technologies, Limited
Made in
United Kingdom
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →