Heartware, Inc. recalls Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers: 1403, 1407, and 1420 Product Usage: The H

Recalled October 10, 2018 · FDA recall Z-0067-2019 · Heartware

Hazard
Certain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the controller housing. · FDA Class II
Units
17,275 units
Sold at
Worldwide Distribution

Description

Certain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the controller housing.

Product
Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers: 1403, 1407, and 1420 Product Usage: The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.
Manufacturer(s)
Heartware, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov