Heartware, Inc. recalls Medtronic HVAD Packaged Monitor, Model Numbers: 1500, 1500AU, 1500DE, 1510US, 1511AR, 1511AU, 1511BR, 1511CA, 1511CH, 1
Recalled April 14, 2021 · FDA recall Z-1341-2021 · Heartware
Hazard
There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port. · FDA Class I
Units
3283 devices (1824 US)
Sold at
Worldwide
Description
There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.
Product
Medtronic HVAD Packaged Monitor, Model Numbers: 1500, 1500AU, 1500DE, 1510US, 1511AR, 1511AU, 1511BR, 1511CA, 1511CH, 1511DE, 1511GB, 1511IL, 1511IN, 1511IT, 1511JP, 1511MY, 1520CLIN-AU, 1520CLIN-DE, 1520CLIN-GB, 1520JP, 1520US, 1521AU, 1521BR, 1521CA, 1521CH, 1521DE, 1521GB, 1521IL, 1521IN, 1521IT, 1521JP, 1521KR, 1521TW, 1521US, 1522CA, 1522CA-CLIN
Manufacturer(s)
Heartware, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →