RecallWatchUSA

Home / Recalls / Medical Devices

Heartware, Inc. recalls Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT

Recalled April 14, 2021 · FDA recall Z-1337-2021 · Heartware

Hazard
There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port. · FDA Class I
Units
17992 units (8054 US)
Sold at
Worldwide

Description

There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.

Product
Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US 2. Medtronic HVAD Controller DC Adapter, Product Number: 1440
Manufacturer(s)
Heartware, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Heartware, Inc. recalls Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT | RecallWatchUSA