Helena Laboratories, Corp. recalls SPIFE Alkaline Hemoglobin Kit, Cat. No. 3415, for use in the separation and quantitation of hemoglobins by agarose e

Recalled November 13, 2019 · FDA recall Z-0297-2020 · Helena Laboratories

Hazard
Lots of gel kits contained a newer revision of the Customer Procedures instructions than intended. The incorrect inserts refer to updated parameters associated with new electrodes -- which have not yet been made available to users. · FDA Class III
Units
899 kits
Sold at
US: OH, NY, MI, IN, IL, MS, TX, CT, NJ, KY, FL, MD, GA, PA, AZ, MO, VA, SC, LA, KS, WA, WI, OUS: Colombia, Brasil, Taiwan

Description

Lots of gel kits contained a newer revision of the Customer Procedures instructions than intended. The incorrect inserts refer to updated parameters associated with new electrodes -- which have not yet been made available to users.

Product
SPIFE Alkaline Hemoglobin Kit, Cat. No. 3415, for use in the separation and quantitation of hemoglobins by agarose electrophoresis using the SPIFE systems
Manufacturer(s)
Helena Laboratories, Corp.