Home / Recalls / Medical Devices
Hill-Rom, Inc. recalls Centrella Smart+ Bed, Catalog Number P7900
Recalled June 10, 2020 · FDA recall Z-2246-2020 · Hill-Rom
Hazard
The Bed Exit System may fail to send a remote alert through the nurse call system if a remote alert was previously sent and cancelled at the in-room nurse call wall unit. · FDA Class II
Units
4381 beds
Sold at
Domestic distribution nationwide. Foreign distribution to Canada and Australia.
Description
The Bed Exit System may fail to send a remote alert through the nurse call system if a remote alert was previously sent and cancelled at the in-room nurse call wall unit.
Product
Centrella Smart+ Bed, Catalog Number P7900
Manufacturer(s)
Hill-Rom, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →