Hiossen Inc. recalls ET Rigid Abutment (Mini) Size: 4.5D 2.0G/H 4.0H- Intended for use with a dental implant to provide support for prostheti

Recalled March 12, 2025 · FDA recall Z-1275-2025 · Hiossen

Hazard
Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments as Regular version whereas the actual product is a Mini version; and identify the ET Multi Abutments as Mini version whereas the actual product is a Regular version · FDA Class II
Units
19 units
Sold at
US Nationwide distribution.

Description

Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments as Regular version whereas the actual product is a Mini version; and identify the ET Multi Abutments as Mini version whereas the actual product is a Regular version

Product
ET Rigid Abutment (Mini) Size: 4.5D 2.0G/H 4.0H- Intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Model/Catalog Number: ETRGA4524MP
Manufacturer(s)
Hiossen Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov