Hologic, Inc. recalls Hologic Fluoroscan Mini C-Arm Model with InSight2, Software version 4.0 and 4.0.1 The device is a Mini C-arm Fluoros

Recalled October 3, 2012 · FDA recall Z-2452-2012 · Hologic

Hazard
InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180¿around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180¿ around vertical axis as well as 180¿ around a horizontal axis. · FDA Class II
Units
414 units
Sold at
Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia,Belgium, Canada, China, Germany, Ecuador, Spain, France, Great Britain, Greece, Hong Kong, Italy, Japan Korea, Lithuvania, Neherlands, qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan, South Africa, and Switzerland.

Description

InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180¿around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180¿ around vertical axis as well as 180¿ around a horizontal axis.

Product
Hologic Fluoroscan Mini C-Arm Model with InSight2, Software version 4.0 and 4.0.1 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.
Manufacturer(s)
Hologic, Inc.