RecallWatchUSA

Home / Recalls / Medical Devices

Hospira Inc. recalls GemStar infusion pump, list numbers 13086 and 13087, GemStar infusion pump (pain management) list number 13088. Prod

Recalled December 3, 2014 · FDA recall Z-0422-2015 · Hospira

Hazard
Some patients are obtaining access to the dosage reset codes required to change the settings on GemStar infusion pumps and are subsequently changing the infusion rate set by their doctors. Patients are obtaining access to these codes via website sponsors by downloading the GemStar User/Service Manual featured on the sponsors' websites. The GemStar User/Service Manual should be removed from these w · FDA Class II
Units
Not Applicable
Sold at
Worldwide Distribution - US: Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Croatia, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Icelanofd, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malta, Netherlands, New Zealand, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom

Description

Some patients are obtaining access to the dosage reset codes required to change the settings on GemStar infusion pumps and are subsequently changing the infusion rate set by their doctors. Patients are obtaining access to these codes via website sponsors by downloading the GemStar User/Service Manual featured on the sponsors' websites. The GemStar User/Service Manual should be removed from these w

Product
GemStar infusion pump, list numbers 13086 and 13087, GemStar infusion pump (pain management) list number 13088. Product Usage The pump is suitable for intravenous, arterial, subcutaneous, short-term epidural infusion and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products. The epidural route is recommended to provide anesthesia or administer analgesia for periods up to 96 hours. ..
Manufacturer(s)
Hospira Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Hospira Inc. recalls GemStar infusion pump, list numbers 13086 and 13087, GemStar infusion pump (pain management) list number 13088. Prod | RecallWatchUSA