Home / Recalls / Medical Devices
Hospira Inc. recalls Hospira RECEPTAL LINERS: 1L(1000mL), 1.5L (1500mL), 2L (2000mL) 1) List Number 43056-01; 2) List Number 43023-01;
Recalled August 24, 2016 · FDA recall Z-2594-2016 · Hospira
Hazard
Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent. · FDA Class II
Units
67,410,861 liners and 67,410,861 lids/canisters in total
Sold at
*** US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY), District of Columbia, Puerto Rico; *** FOREIGN: Australia, Canada, Costa Rica, Singapore, South Africa
Description
Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.
Product
Hospira RECEPTAL LINERS: 1L(1000mL), 1.5L (1500mL), 2L (2000mL) 1) List Number 43056-01; 2) List Number 43023-01; 3) List Number 43025-01; 4) List Number 43025-11; 5) List Number 43025-21; 6) List Number 43025-25; 7) List Number 43038-05; 8) List Number 43043-01; 9) List Number 43044-01; 10) List Number 43044-11; 11) List Number 43046-01; 12) List Number 43047-01; 13) List Number 43041-01; 14) List Number 43024-01; 15) List Number 43027-01; 16) List Number 43027-14; 17) List Number 43001-01; 18) List Number 43064-01; 19) List Number 43046-05; 20) List Number 43042-11; 21) List Number 43042-01; 22) List Number 43044-05; 23) List Number 43044-15
Manufacturer(s)
Hospira Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →