Hazard
Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion. · FDA Class II
Units
38,878 infusion pump systems
Sold at
Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY) + District of Columbia + Puerto Rico + Virgin Islands; *** FOREIGN:

Description

Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.

Product
LifeCare PCA infusion system with Hospira MedNet software allows clinicians to administer, or patients to self-administer analgesia, safely and effectively within clinician programmed limits and/or hospital-defined medication limits. The LifeCare PCA infusion system is used in a wide range of clinical settings that includes but is not limited to the following: medical, laboratory/delivery/post-partum, burn unit, surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), pediatrics.
Manufacturer(s)
Hospira Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Hospira Inc. recalls LifeCare PCA infusion system with Hospira MedNet software allows clinicians to administer, or patients to self-administe | RecallWatchUSA