Home / Recalls / Medical Devices
HUMAN MED AG recalls human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use.
Recalled March 12, 2025 · FDA recall Z-1251-2025 · HUMAN MED AG
Hazard
The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet. · FDA Class II
Units
775 devices
Sold at
US Nationwide distribution in the state of FL.
Description
The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet.
Product
human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use.
Manufacturer(s)
HUMAN MED AG
Made in
Germany
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →