RecallWatchUSA

Home / Recalls / Medical Devices

I.T.S. GmbH recalls FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.

Recalled April 1, 2026 · FDA recall Z-1595-2026 · I.T.S

Hazard
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). · FDA Class II
Units
4 units
Sold at
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Description

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Product
FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.
Manufacturer(s)
I.T.S. GmbH
Made in
Austria

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

I.T.S. GmbH recalls FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60. | RecallWatchUSA