I.T.S. GmbH recalls Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hoo
Recalled April 1, 2026 · FDA recall Z-1597-2026 · I.T.S
Hazard
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). · FDA Class II
Units
70 units
Sold at
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Description
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Product
Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12.
Manufacturer(s)
I.T.S. GmbH
Made in
Austria
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →