RecallWatchUSA

Home / Recalls / Medical Devices

I.T.S. GmbH recalls Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral T

Recalled April 1, 2026 · FDA recall Z-1596-2026 · I.T.S

Hazard
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). · FDA Class II
Units
101 units
Sold at
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Description

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Product
Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 21094-8. 3. Pilon Plate, Small; Article Numbers: 21093-4, 21093-6.
Manufacturer(s)
I.T.S. GmbH
Made in
Austria

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

I.T.S. GmbH recalls Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral T | RecallWatchUSA