Home / Recalls / Medical Devices
Iba Dosimetry recalls Large Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T, DS50-103-XL-T
Recalled June 14, 2017 · FDA recall Z-2220-2017 · Iba Dosimetry
Hazard
When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear. · FDA Class II
Units
110 units
Sold at
Worldwide Distribution - USA (nationwide) and to the countries of : Argentina, Australia, Austria, Canada, Cyprus, France, Germany, Hong Kong, Ireland, Islamic Republic of, Israel, Italy, Japan, Kingdom of Saudi Arabia, Kuwait, New Zealand, People's Republic of China, Philippines, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, R.O.C., and United Kingdom.
Description
When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear.
Product
Large Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T, DS50-103-XL-T
Manufacturer(s)
Iba Dosimetry
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →