Home / Recalls / Medical Devices
ICU Medical de Mexico, S.A. de C.V. recalls 191 cm (75") Appx 2.9 ml, PVC/PUR Smallbore Bifuse Ext Set w/6-Port NanoClave Manifold w/2 Red Rings, Check Valve, 2 Mic
Recalled August 22, 2018 · FDA recall Z-2807-2018 · ICU Medical de Mexico, S.A. de C.V
Hazard
ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold. · FDA Class II
Units
50 pouches
Sold at
US distribution to AZ, MA, NC, OH, TX, VA, WA. Worldwide distribution to Belgium, France, Germany.
Description
ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold.
Product
191 cm (75") Appx 2.9 ml, PVC/PUR Smallbore Bifuse Ext Set w/6-Port NanoClave Manifold w/2 Red Rings, Check Valve, 2 MicroClave Clear, Clamp, Rotating Luer.
Manufacturer(s)
ICU Medical de Mexico, S.A. de C.V.
Made in
Mexico
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →