ICU Medical, Inc. recalls ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular Administra
Recalled October 10, 2018 · FDA recall Z-0069-2019 · ICU Medical
Hazard
Design change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product design change and use the transfer set as a blood transfusion set without realizing a blood filter is not included. · FDA Class II
Units
2361 Units
Sold at
U.S. Distribution in states of: NY, AL, CA, MS, PA, GA, MI, NC, WI, IL, WA, UT and FL.
Description
Design change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product design change and use the transfer set as a blood transfusion set without realizing a blood filter is not included.
Product
ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Manufacturer(s)
ICU Medical, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →