RecallWatchUSA

Home / Recalls / Medical Devices

ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

Recalled December 31, 2025 · FDA recall Z-0966-2026 · ICU Medical

Hazard
Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery. · FDA Class II
Units
12 customers
Sold at
US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.

Description

Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.

Product
LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02
Manufacturer(s)
ICU Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov