Home / Recalls / Medical Devices
ICU Medical, Inc. recalls QUAD KIT WITH THREE TRANSPAC¿ AND 03ML FLUSH DEVICES FOR ST. FRANCIS HOSP., Item No. 46094-47 The Transpac Disposable S
Recalled November 23, 2016 · FDA recall Z-0513-2017 · ICU Medical
Hazard
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers. · FDA Class II
Units
200 units
Sold at
Nationwide Distribution
Description
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
Product
QUAD KIT WITH THREE TRANSPAC¿ AND 03ML FLUSH DEVICES FOR ST. FRANCIS HOSP., Item No. 46094-47 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a compatible pressure monitor. The SafeSet Reservoir system is a blood sampling system assembled directly to the patient monitoring line (Transpac units). The SafeSet reservoir is used on patients where periodic blood sampling is required.
Manufacturer(s)
ICU Medical, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →