Home / Recalls / Medical Devices
IHB OPERATIONS B.V. recalls Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
Recalled April 30, 2025 · FDA recall Z-1663-2025 · IHB OPERATIONS B.V
Hazard
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column. · FDA Class II
Units
112 units
Sold at
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
Description
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Product
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
Manufacturer(s)
IHB OPERATIONS B.V.
Made in
Netherlands
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →