RecallWatchUSA

Home / Recalls / Medical Devices

Immuno-Mycologics, Inc recalls Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003

Recalled February 8, 2017 · FDA recall Z-1109-2017 · Immuno-Mycologics

Hazard
The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results. · FDA Class II
Units
630 units
Sold at
Nationwide throughout the US

Description

The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.

Product
Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003
Manufacturer(s)
Immuno-Mycologics, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov