Immuno-Mycologics, Inc recalls CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and
Recalled February 9, 2022 · FDA recall Z-0545-2022 · Immuno-Mycologics
Hazard
The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected. · FDA Class II
Units
411 kits
Sold at
US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA.
Description
The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.
Product
CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and CSF, REF CR2003, IVD, UDI: (01)00816387020919
Manufacturer(s)
Immuno-Mycologics, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →