Home / Recalls / Medical Devices
IMPLANTCAST GMBH recalls EcoFit cup cementless: Model(Catalog Number)/Description: 02200046 /EcoFit cup cementless 46mm dia incl. central hole
Recalled July 21, 2021 · FDA recall Z-2050-2021 · IMPLANTCAST
Hazard
A risk exists during impacting of the acetabular cup into the bone the plugs may come loose from the cup, which could lead to operation duration extension or revision operation. A warning note is being added to instructions for use and surgical technique alerting users that the quality of bone in the patient (such as osteosclerosis) is an attributable factor. · FDA Class II
Units
41 units (U.S.)
Sold at
U.S. Nationwide distribution in the states of FL and TX.
Description
A risk exists during impacting of the acetabular cup into the bone the plugs may come loose from the cup, which could lead to operation duration extension or revision operation. A warning note is being added to instructions for use and surgical technique alerting users that the quality of bone in the patient (such as osteosclerosis) is an attributable factor.
Product
EcoFit cup cementless: Model(Catalog Number)/Description: 02200046 /EcoFit cup cementless 46mm dia incl. central hole cover; 02200048 /EcoFit cup cementless 48mm dia incl. central hole cover; 02200050 /EcoFit cup cementless 50mm dia incl. central hole cover; 02200052 /EcoFit cup cementless 52mm dia incl. central hole cover; 02200054 /EcoFit cup cementless 54mm dia incl. central hole cover; 02200056 /EcoFit cup cementless 56mm dia incl. central hole cover; 02200058 /EcoFit cup cementless 58mm dia incl. central hole cover; 02200060 /EcoFit cup cementless 60mm dia incl. central hole cover; 02200062 /EcoFit cup cementless 62mm dia incl. central hole cover; 02200064 /EcoFit cup cementless 64mm dia incl. central hole cover; 02200066 /EcoFit cup cementless 66mm dia incl. central hole cover; 02200068 /EcoFit cup cementless 68mm dia incl. central hole cover;
Manufacturer(s)
IMPLANTCAST GMBH
Made in
Germany