IMPRIS Imaging Inc recalls IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida

Recalled November 5, 2025 · FDA recall Z-0164-2026 · IMPRIS Imaging

Hazard
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room. · FDA Class I
Units
N/A
Sold at
Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.

Description

There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Product
IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida
Manufacturer(s)
IMPRIS Imaging Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov