Infusion Pump Repair recalls Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part
Recalled May 19, 2021 · FDA recall Z-1544-2021 · Infusion Pump Repair
Hazard
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion. · FDA Class I
Units
15
Sold at
California
Description
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Product
Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part
Manufacturer(s)
Infusion Pump Repair
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →