Home / Recalls / Medical Devices
Innovasis, Inc recalls Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System
Recalled August 12, 2026 · FDA recall Z-2859-2026 · Innovasis
Hazard
Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step. · FDA Class II
Units
5
Sold at
US Nationwide distribution in the states of TX, CA.
Description
Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.
Product
Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System
Manufacturer(s)
Innovasis, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →