RecallWatchUSA

Home / Recalls / Medical Devices

Innovasis, Inc recalls Oryx Cervical Screw Caddy, Model Number BTA-172, as part of the Oryx Cervical Plate System - Product Usage: The Cervical

Recalled May 27, 2020 · FDA recall Z-2062-2020 · Innovasis

Hazard
Cervical screw caddies have incorrect text markings for the 4.5mm Fixed and Variable rescue screws. The caddy location labeled as "Fixed" should be labeled as "Variable" and the caddy location labeled as "Variable" should be labeled as "Fixed. If user installs a fixed rescue screw instead of variable, the screw head may not be flush, it could back out, which could interfere with the cover screw. · FDA Class II
Units
6
Sold at
US Nationwide distribution including in the states of CA, AL, LA.

Description

Cervical screw caddies have incorrect text markings for the 4.5mm Fixed and Variable rescue screws. The caddy location labeled as "Fixed" should be labeled as "Variable" and the caddy location labeled as "Variable" should be labeled as "Fixed. If user installs a fixed rescue screw instead of variable, the screw head may not be flush, it could back out, which could interfere with the cover screw.

Product
Oryx Cervical Screw Caddy, Model Number BTA-172, as part of the Oryx Cervical Plate System - Product Usage: The Cervical Plate System is intended for use in anterior cervical fixation.
Manufacturer(s)
Innovasis, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov